[. . . ] Thompson User Manual Panniculectomy Retractor System
jckdpanum0307
1
Opposite the surgeon, at the patient's nipple line, secure the Elite II Rail Clamp (#41902AC) over the sterile drape to the OR table.
2
Insert the Collapsible Grooved Rod (#44734) into
the Elite II Rail Clamp's cam joint with the grooved portion in the horizontal position. [. . . ] Lock into place by actuating the handle and pressing it tight.
5
Attach Perforating Towel Clips to pannus as needed and loop the finger rings over the Towel Clip Crank hooks.
Retract as needed by turning the Towel Clip Crank's knobs.
jckdpanum0307
C A R E & MAINTENANCE
CLEAN ALL COMPONENTS. Ultrasonic cleaning is recommended; however, hand washing with a brush is acceptable. Be certain that all Cam Joints are in FULL OPEN POSITION during autoclaving. STERILIZE: GRAVITY STEAM (Wrapped) 30 minutes 121°C (250°F) @ 1. 1kg/cm2 HIGH SPEED GRAVITY (Flash) 5 minutes 132°C (270°F) @30psi or 132° @ 2kg/cm2 PREVACUUM STEAM (Wrapped) 10 minutes @ 270°F or 132°C These sterilization settings are recommendations. Consult your autoclave manual for specific instructions, conditions and exposure periods. FOLLOWING STERILIZATION, ALLOW YOUR INSTRUMENTS TO COOL BEFORE HANDLING. [. . . ] Elite II Rail Clamp with Cam Joint 41902AC Towel Clip Cranks 35028
10170 East Cherry Bend Road Traverse City, Michigan 49684 (800) 227-7543 | Fax (231) 922-0174 For Customer Service call (800) 227-7543 or (231) 922-0177 Or Visit us on the World Wide Web at
Straight Arm with 1 Cam Joint 44120C
www. thompsonsurgical. com
90° Collapsible Rod with Grooves 44734
jckdpanum0307 [. . . ]